Counting Cells Using a Hemacytometer

1.         Do not touch the face of the coverslip or the hemacytometer face with your fingers. Grease from you fingers can prevent the chamber from filling properly.

 

2.         Rinse the hemacytometer and coverslip with distilled water, dry using kimwipes and clean with lens paper.

 

3.         Mix sample with Pasteur pipette.  With the coverslip in place, use a Pasteur pipette to transfer a small amount of the cell suspension to the counting chamber. Carefully touch the edge of the coverslip with the pipette tip and allow the chamber to fill by capillary action. Repeat with other chamber, mixing sample well before loading.

 

4.         Count cells in the four corner squares and the middle square (see diagram below)

 

5.         Each square of the hemacytometer (with cover slip in place) represents a volume of 0.1 mm3 or 10-4 cm3. Since 1 cm3 is equivalent to 1 mL, the subsequent cell concentration per mL (and the total number of cells) will be determined using the following calculations.

 

         Cells per mL= the average number of cells per square x 104

Example: If the average count per square is 45 cells, the density of cells is 45 x 104 cell mL-1

If the sample was diluted prior to counting, you should multiple the number above by the dilution factor

Total cell number = cells per mL x the original volume of fluid from which cell sample was removed.

Example: 4.5 x 10cells mL-1 x 10 mL (culture volume) =4.5 x 106 cells total in the culture volume.

 

6.         Clean the hemacytometer with distilled water, dry, and place in storage box.

 

Reference: Sigma Catalog (2000-2001) page 1849